November 04, 2025

Get In Touch

Novel Electronic Pulp Sensitivity Tester With Transilluminator Efficient And Safe Method To Heat Test Pulp

Study on Electronic Pulp Sensitivity Tester

A Novel Electronic Pulp Sensitivity Tester with a Transilluminator

An efficient and safe method to heat test pulp suggests a new study published in the Journal of Endodontics.

Study Overview

The efficacy and safety of a single novel electronic pulp sensitivity tester with a transilluminator (PSTT) capable of providing 4 different tests were compared with gold standard (GS) pulp testing methods.

Four hundred eighty teeth, including 3 from each quadrant and their contralateral and opposing teeth, were randomly assessed using the PSTT or GS methods. Seven days later, the same teeth were assessed using the method not used earlier so that all teeth were evaluated using both methods. Sixty previously root canal–treated teeth, serving as negative controls, were assessed identically to the experimental groups. Results were analyzed using IBM SPSS software (IBM Corp, Armonk, NY) (P < .05).

Results

A higher percentage of teeth responded to cold and electricity using GS methods compared with the PSTT (99.4 vs 93.1 and 99.6 vs 97.3, respectively). A lower percentage of teeth responded to heat using the GS method compared with the PSTT (50.0 vs 68.1). Chi-square tests determined differences in the proportion of sensitivity to temperature, and electric testing methods were statistically significant (P < .001 and P = .004, respectively). No difference in the proportion of sensitivity was observed for transillumination. The Wilcoxon signed rank test determined significantly shorter cold testing times using the GS method (P = .024). Shorter testing times were observed using the PSTT for heat (P < .001), electric pulp testing (P = .048), and transillumination (P = .001). The overall PSTT testing time was significantly shorter than the GS testing time (P = .03). Tissue injury was not observed.

The PSTT efficiently and safely provided heat and transillumination for pulpal diagnosis. Improvements are needed to enhance the cold and electric stimulus efficacy.

Reference

Comparison of the Efficacy and Safety of Pulp Sensitivity Testing Results Using a New Pulp Testing Device versus Some Commonly Used Methods: A Crossover Study

Authors: Patricia A. Tordik, Mahmoud Torabinejad, Cassandra M. Fevelo, Mason Bahador, DDS, MS, Udochukwu Oyoyo

Published: October 13, 2023

DOI: https://doi.org/10.1016/j.joen.2023.10.003

Keywords

  • Novel
  • Electronic
  • Pulp
  • Sensitivity
  • Tester
  • Transilluminator
  • Efficient
  • Safe
  • Method
  • Heat
  • Test
  • Pulp
  • Testing Device versus Some Commonly Used Methods: A Crossover Study
  • Patricia A. Tordik
  • Mahmoud Torabinejad
  • Cassandra M. Fevelo
  • Mason Bahador
  • DDS
  • MS
  • Udochukwu Oyoyo

Disclaimer: This website is designed for healthcare professionals and serves solely for informational purposes.
The content provided should not be interpreted as medical advice, diagnosis, treatment recommendations, prescriptions, or endorsements of specific medical practices. It is not a replacement for professional medical consultation or the expertise of a licensed healthcare provider.
Given the ever-evolving nature of medical science, we strive to keep our information accurate and up to date. However, we do not guarantee the completeness or accuracy of the content.
If you come across any inconsistencies, please reach out to us at admin@doctornewsdaily.com.
We do not support or endorse medical opinions, treatments, or recommendations that contradict the advice of qualified healthcare professionals.
By using this website, you agree to our Terms of Use, Privacy Policy, and Advertisement Policy.
For further details, please review our Full Disclaimer.

0 Comments

Post a comment

Please login to post a comment.

No comments yet. Be the first to comment!