High-Speed Dental Handpiece With Coolant Delivery Limits Aerosol Contamination

Study on Electric Micromotor Handpieces

Study on Electric Micromotor Handpieces

Electric micromotor handpieces which use water-jet coolant alone without compressed air produce localized (within treatment bay) droplet contamination but are unlikely to produce aerosol contamination beyond the immediate treatment area, finds a study.

The study is published in the Journal of Dentistry.

High-speed dental instruments produce aerosol and droplets. James R Allison et al from the School of Dental Sciences, Faculty of Medical Sciences, Newcastle University, United Kingdom carried out the present study with the objective to evaluate aerosol and droplet production from a novel electric micromotor handpiece (without compressed air coolant) in real world clinical settings.

The authors performed 10-minute upper incisor crown preparations in triplicate in an open-plan clinic with mechanical ventilation providing 3.45 air changes per hour. A 1:5 ratio electric micromotor handpiece which allows water coolant without compressed air (Ti-Max Z95L, NSK) was used at three speeds: 60,000 (60 K), 120,000 (120 K), and 200,000 (200 K) revolutions per minute.

Coolant solutions contained fluorescein sodium as a tracer (2.65 mmol L − 1). High-speed air-turbine positive control, and negative control conditions were conducted. Aerosol production was evaluated at 3 locations (0.5 m, 1.5 m, and 1.7 m) using:

  • An optical particle counter (OPC; 3016-IAQ, Lighthouse) to detect all aerosol
  • A liquid cyclone air sampler (BioSampler, SKC Ltd.) to detect aerosolised fluorescein, which was quantified by spectrofluorometric analysis

Settled droplets were detected by spectrofluorometric analysis of filter papers placed onto a rig across the open-plan clinic in the study.

Findings

  • Local (within treatment bay) settled droplet contamination was elevated above negative control for all conditions, with no difference between conditions.
  • Settled droplet contamination was not detected above negative controls outside the treatment bay for any condition.
  • Aerosol detection at 1.5 m and 1.7 m, was only increased for the air-turbine positive control condition.
  • At 0.5 m, aerosol levels were highly elevated for the air-turbine, minimally elevated for 200 K and 120 K, and not elevated for 60 K.

Hence, the authors concluded that "electric micromotor handpieces which use water-jet coolant alone without compressed air produce localized (within treatment bay) droplet contamination, but are unlikely to produce aerosol contamination beyond the immediate treatment area (1.5 m), allowing them to be used safely in most open-plan clinic settings."

Disclaimer: This website is designed for healthcare professionals and serves solely for informational purposes.
The content provided should not be interpreted as medical advice, diagnosis, treatment recommendations, prescriptions, or endorsements of specific medical practices. It is not a replacement for professional medical consultation or the expertise of a licensed healthcare provider.
Given the ever-evolving nature of medical science, we strive to keep our information accurate and up to date. However, we do not guarantee the completeness or accuracy of the content.
If you come across any inconsistencies, please reach out to us at admin@doctornewsdaily.com.
We do not support or endorse medical opinions, treatments, or recommendations that contradict the advice of qualified healthcare professionals.
By using this website, you agree to our Terms of Use, Privacy Policy, and Advertisement Policy.
For further details, please review our Full Disclaimer.

0 Comments

Post a comment

Please login to post a comment.

No comments yet. Be the first to comment!